BC Biomedical DPM-2100 Uživatelský manuál Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 28
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
14
PRESSURE
RANGE
-13.50 to 100.00 PSI
-701 to 5190 mmHg @ 20 °C
-374 to 2773 inH
2
O @ 20 °C
-951 to 7043 cmH
2
O @ 20 °C
RESOLUTION
0.01 PSI
1 mmHg @ 20 °C
1 inH
2
O @ 20 °C
1 cmH
2
O @ 20 °C
MAXIMUM
OVERPRESSURE
200 PSI
ACCURACY
± 0.1% FS
DIGITAL FILTER
1 255 samples
COMPATIBLE MEDIA
Only non-corrosive, non-ionic, or otherwise pure fluids
and/or gases that are compatible with sensor materials
including glass, silicon, ceramic, epoxy, RTV, gold,
aluminum and nickel.
CONNECTIONS
Male Luer Coupler
TEMPERATURE (DPM-2100 ONLY)
RANGE
0.0 to 100.0 °C (32.0 to 212.0 °F)
RESOLUTION
0.1 °C (0.1 °F)
ACCURACY
± 0.5% FS
CONNECTIONS
¼ inch Phone Jack for ¼ Phone Plug terminated
temperature cables or probes
SENSOR
COMPATIBILITY
YSI Series 700 Temperature Probes or equivalent
PHYSICAL
DISPLAY
5 Digit, 7-Segment Non-Backlit LCD
CONSTRUCTION
ENCLOSURE
ABS Plastic
OVERLAY
Lexan, Back Printed
MEMORY
SETUP
EEPROM, All Parameters
RETENTION
10 Years w/o Power
SIZE
7.69 x 3.97 x 1.82 Inches (195.3 x 100.8 x 46.2 mm)
WEIGHT
≤ 1 Lbs (0.45 kg)
SPECIFICATIONS
Zobrazit stránku 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 28

Komentáře k této Příručce

Žádné komentáře