The following is a breakdown of the parameters available in the configuration of the unit
and their available options:
System Setup Configuration
Selects the RF Current Transformer Attenuation in
Volts : Amp for the RF Donut being used.
Default = 0.1 : 1
0.1 : 1
1 : 1
Volts : Amps
Zeros the input circuitry based on donut being
used. Each donut can have a slightly different zero
offset. This function will eliminate the offset from
the readings. Press the SETUP key while this
parameter is selected to perform the auto –zero
function.
Determines the input range.
Default = Auto
Input Mode
(ESU-2050P Only)
Determines whether the unit continuously monitors
the Input Signal or looks for a Pulsed Input Signal.
Default = Continuous
Used for Power (Watts) calculation only. Can be
adjustable or set by a table of fixed resistors. (See
Load Selection and Load Table for more
information.)
Default = 500
Determines whether the Load Resistance
Parameter is adjustable by tenths of ohms or
selected from the Load Table. The Load Table is
created from the Load Resistance Values set in the
Factory Setup and the Load Sets. (See Load Sets,
Load Table and Factory Setup for more
information.)
Default = Table
Determines the power up mode of the ESU-2050.
The default mode shows a single mV parameter
display. Set this parameter to custom to display
the saved startup mode. Set this parameter to
Save current as custom to save the settings for the
next time power is cycled.
Default = Defaults
Defaults
Custom
Set Current as Custom
Sets the number of A/D converter readings used in
each mV RMS computation. A higher setting
requires more computation and is slower, but
results in a more stable reading.
Default = 32,768
1024
2048
4096
8192
16384
32768
Sets which display averaging parameter is to be
used. Three independent averaging modes can be
configured for optimum system performance.