System Setup Configuration
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter is
set to Slow. A higher number will cause the display
to update slower, but will give a more stable
reading
Default = 150
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter is
set to Medium. A higher number will cause the
display to update slower, but will give a more stable
reading
Default = 15
Sets the number of mV RMS readings that are
averaged when the Display Averaging parameter is
set to Fast. A higher number will cause the display
to update slower, but will give a more stable
reading
Default = 4
Sets the range of input readings that will be
averaged. If a new mV reading deviates from the
average by less than this amount, it will be
averaged with the rest of the readings in the
display averaging buffer. Otherwise, the input is
considered a step change and the display
averaging buffer is flushed.
Assigns a resistance value for Set 1 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
Assigns a resistance value for Set 2 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
Assigns a resistance value for Set 3 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected
Assigns a resistance value for Set 4 from a
combination of the available loads as determined
by the Load Resistance Values set in the Factory
Setup. (See Custom Load Sets for more
information.)
Default = None Selected