4-4
GETTING READY FOR A DISTAL OCCLUSION TEST
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--OCCLYZER--OCCL
YZER--OCCLYZER--OCCL
YZER--OCCL
USIONUSION
USIONUSION
USION
Choose Channel Screen
To run a test, toggle to the appropriate Channel screen
using .
STEP 2
9
Connect and Prime Device
Connect device under test (DUT) to the unit and prime the
channel. (See PRIMING for details.)
There is an optional NURSE CALL input where the NURSE
CALL output from the DUT can be connected to allow the
unit to record when the DUT sends a NURSE CALL
message.
STEP 1
9
GETTING READY FOR A DISTAL OCCLUSION TEST
NOTES:
C
The NORMAL LED on the remote
module will blink at a rate of twice per
second, when a channel needs to be
primed.
C
As the channel becomes ready, the
NORMAL LED on the remote module
will blink at a rate of once per second.
C
The priming message will not display
if the DUT is already primed.
NOTES:
C
There can be up to two modules
connected to the analyzer.
C
The modules are interchangeable and
the unit will run with a single module
connected to either port.
C
The Channel 1 or 2 screens will only
be displayed when the modules are
connected to those ports.
C
These instructions will utilize Channel
1 for examples; Channel 2 functions
identically.
CHANNEL 1 IDLE
PRIME WITH 50 mL
THEN PRESS START
Zobrazit stránku 113