NOTES:
C
This sample shows both summary
and data information.
C
There are three options for the form
of the report: Summary, Data Points
or Summary and Data.
C
This option can be selected from the
Datalog screen.
C(See DATALOG--SAFETY
ANALYZER for details.)
PAGE 01
BC GROUP
IPA-2000 INFUSION PUMP ANALYZER
PUMP INFORMATION:
MANUFACTURER: PUMP MAKER MODEL: 123
SERIAL NUMBER: 456789
DEPARTMENT: ER LOCATION: HOSPITAL
TEST STARTED AT: 14:59:01 ON 6/27/02
SAFETY ANALYZER DATA
LINE VOLTAGE 125.3V
GROUND RESISTANCE .08ohm
LOAD CURRENT .06A
NORMAL GROUND CURRENT 15uA
MEASUREMENT METHOD
DUT: ON POL: FWD
NEU:CLOSE HOT:CLOSE
FAULT GROUND CURRENT 25uA
MEASUREMENT METHOD
DUT: ON POL: REV
NEU: OPEN HOT:CLOSE
NORMAL CHASSIS CURRENT 02uA
MEASUREMENT METHOD
DUT: OFF POL: REV
GND:CLOSE
FAULT CHASSIS CURRENT 15uA
MEASUREMENT METHOD
DUT: OFF POL: FWD
GND: OPEN
SAFETY ANALYZER SUMMARY
PASS - WITHIN LIMITS
SAFETY SPECS AAMI
MAX GROUND RESISTANCE .15ohm
MAX NORMAL GROUND CURRENT 500uA
MAX FAULT GROUND CURRENT 1000uA
MAX NORMAL CHASSIS CURRENT 100uA
MAX FAULT CHASSIS CURRENT 300uA
PASS / FAIL REPORT MESSAGES
L
NOTE:
CThe Summary reports a Pass or
Fail message, telling which state
caused the failure.
PASS - WITHIN LIMITS
FAIL - PLUG POLARITY
FAIL - LINE VOLTAGE
FAIL - LOAD CURRENT
FAIL - OPEN FUSE
FAIL - NORM GND CUR
FAIL - FAULT GND CUR
FAIL - NORM CHS CUR
FAIL - FAULT CHS CUR