PART 3: INFUSION ANALYZER--FLOW
There are two basic types of flow tests that may be run:
Flow and Flow--PCA.
In the standard Flow test, a continuous flow is assumed and
the data is recorded on a continuous time basis.
In the Flow--PCA test, the system monitors for Bolus flow. It
records the flow verses time, but calculates the average flow
only during the Bolus period. It also records the lockout time
between flows.
A tracking algorithm allows Bolus times with pulsatile
segments to be handled, by analyzing each sub-pulse and
adjusting the test timing accordingly.
The system can also pulse the Patient Control during the
lockout to confirm operation and lockout performance.
Included are:
CAn overview and step by step instructions for
running a Flow test
CAn overview and step by step instructions for
running a Flow--PCA test
CDetails for using the Datalog