4-8
NOTES:
CThe pressure is recorded once per
second, however, it is monitored
every 1/100 of a second. Thus, the
peak pressure may occur between
the stored Data Points.
CThe last Data Point shows the
maximum pressure detected and
the time closest to that reading.
DATALOG SAMPLE FOR A DISTAL OCCLUSION TEST
DATALOG SAMPLE FOR A DISTAL OCCLUSION TEST
NOTES:
C
This sample shows both summary
and data information.
C
There are three options for the form
of the report: Summary, Data Points
or Summary and Data.
C
This option can be selected from the
Datalog screen.
C(See DATALOG--OCCLUSION
TEST for details.)
INFUSION INFUSION
INFUSION INFUSION
INFUSION
ANAN
ANAN
AN
ALAL
ALAL
AL
YZER--OCCLYZER--OCCL
YZER--OCCLYZER--OCCL
YZER--OCCL
USIONUSION
USIONUSION
USION
PAGE 01
BC GROUP
IPA-2000S INFUSION PUMP ANALYZER
PUMP INFORMATION:
MANUFACTURER: ABC MODEL: 123
SERIAL NUMBER: 1001
DEPARTMENT: SURGERY LOCATION: LOCATION 1
TEST STARTED AT: 17:07:00 ON 7/16/02
CHANNEL 1 OCCLUSION PRESSURE TEST
ELAPSED PRESSURE
TIME PSI
00:00 5.7
00:01 5.7
00:02 5.7
00:03 5.7
00:04 5.8
00:05 6.6
00:06 6.8
00:07 6.8
00:08 6.8
00:09 6.8
00:10 6.8
00:11 6.6
00:12 6.6
00:13 6.7
00:14 7.1
00:15 7.1
00:16 7.1
NURSE CALL INPUT
00:17 7.1
00:18 6.8
00:19 6.6
00:14 7.2
END OF DATA
CHANNEL 1 OCCLUSION TEST SUMMARY
MAXIMUM OCCLUSION PRESSURE 7.2PSI
OCCURRED AT TEST TIME 0:00:14
NURSE CALL INPUT AT TIME 0:00:16